首都医科大学学报 ›› 2005, Vol. 26 ›› Issue (4): 505-508.

• 论著·临床研究 • 上一篇    下一篇

组织调整材料的操作时间及固化时间的测量方法

赵晶1, 洪光2, 李英爱2, 村田比吕司2, 滨田泰三2, 金辰3   

  1. 1. 中日友好医院口腔科;2. 广岛大学大学院医齿药学综合研究科齿科补缀学研究室;3. 德山齿科材料有限公司
  • 收稿日期:2005-01-15 修回日期:1900-01-01 出版日期:2005-08-24 发布日期:2005-08-24

The Development of the Measuring Methods of the Working Time and Setting Time of the Tissue Conditioners

Zhao Jing1, Hong Guang2, Li Yingai2, Murata Hiroshi2, Hamada Taizo2, Jin Chen3   

  1. 1. Department of Stomatology, China-Japan Friendship Hospital;2. Department of Prosthetic Dentistry, Graduate School of Biomedical Sciences, Hiroshima University;3. Tokuyama Dental Corporation
  • Received:2005-01-15 Revised:1900-01-01 Online:2005-08-24 Published:2005-08-24

摘要:

目的 利用移位式流变仪(Displacement rheometer),开发适当的测量组织调整材料的操作时间和固化时间的方法。方法 使用5种组织调整材料.每个材料在23℃和37℃条件下各测量3次.在23℃下发生最初弹性回复的时间定义为操作时间,在37℃下弹性复原率表现为稳定数据的时间定义为固化时间。结果 各个组织调整材料的操作时间和固化时间有显著性差异,显示出不同的固化行为.5种组织调整材料中SR的操作时间最短,为(2.7±0.2)min,VG的操作时间最长,为(15.7±0.5)min.VG的固化时间最长[(25.0±0.8)min],SR的固化时间最短[(7.7±1.9)min]。结论 移位式流变仪方法可以用于测量组织调整材料的操作时间和固化时间,并对分析组织调整材料的固化过程有重大意义.

关键词: 组织调整材料, 操作时间, 固化时间, 移位式流变仪

Abstract:

Objective To develop and establish the testing methods appropriate for the determination of the working time and setting time of these materials using the Displacement rheometer.Methods Five commercial tissue conditioners were used. Three tests were carried out for each material both at 23 ℃(working time) and at 37 ℃(setting time). The working time was defined as the time when initial elastic recovery of the material was observed at 23 ℃. The setting time was the time when the elastic recovery reached a final constant value at 37 ℃.Results The significant differences were found among the working time and setting time among the different materials. The setting behavior of each material was different. The SR sample exhibited shortest value of working time with (2.7±0.2)min. And the VG sample exhibited longest value with (15.7±0.5)min. In the setting time, the VG sample exhibited longest value ((25.0±0.8)min), and SR sample exhibited shortest value ((7.7±1.9)min).Conclusion The results suggest that the Displacement Rheometer suits with determination of the working time and setting time of tissue conditioner. The method is useful for the evaluation of the setting behavior of these material.

Key words: tissue conditioner, working time, setting time, displacement rheometer